Not every instrument in a factory needs the same treatment. A sound calibration programme sorts equipment into three groups.
Group 1: Reference and critical instruments
These include:
- master gauges, reference weights, reference thermometers and other standards used to check other instruments
- instruments whose readings decide product release, safety or legal compliance
- instruments named in customer contracts, GMP requirements or regulations
ISO 9001 itself requires traceability. The accredited calibration is the simplest way to demonstrate traceability: the laboratory’s competence has already been assessed by EGAC. Regulators, notified bodies and many multinational customers expect it. For laboratories that are themselves accredited to ISO/IEC 17025, ILAC P10 makes accredited calibration the normal route.
Group 2: Working instruments
Many organisations calibrate their working instruments internally against reference instruments that have accredited certificates. This is legitimate and economical, provided that:
- a written procedure defines the method, points and acceptance criteria
- the reference instrument’s uncertainty is suitable for the tolerance being checked
- the environment is controlled where it matters
- staff are trained and their competence is recorded
- records show the instrument ID, date, reference used, results, uncertainty or acceptance criteria, and who performed the work
Group 3: Indication-only instruments
Instruments never used to make a product decision may not need calibration at all. Label these clearly “for indication only” so they are never used to make a product decision.
Accreditation status for each instrument and range is confirmed by TEQ at quotation stage. The full accredited parameters and ranges are listed in TEQ’s current EGAC scope schedule. TEQ accreditation.